• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION POLYFLUX 17L; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION POLYFLUX 17L; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 109650
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Dyspnea (1816); Diaphoresis (2452)
Event Date 10/19/2021
Event Type  Injury  
Manufacturer Narrative
Initial reporter facility name: (b)(6) hospital.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that approximately 20 minutes after starting treatment with a polyflux 17l, the patient experienced chest tightness, shortness of breath, and sweating.Treatment was discontinued.The patient was treated with oxygen, 50% glucose water 20ml+ calcium gluconate 2g intravenous (iv) infusion and iv infusion of dexamethasone it was reported symptoms were relieved "about 10 minutes later." no additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYFLUX 17L
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
holger - crafoord - strasse 26
hechingen D-723 79
GM   D-72379
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key12837111
MDR Text Key284628471
Report Number9611369-2021-00308
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414086626
UDI-Public(01)07332414086626
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number109650
Device Lot Number1-4205-H-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-