| Model Number |
6500 |
| Medical Device Problem Codes |
Device Alarm System (1012); Connection Problem (2900); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received indicating that during use of this smiths medical cadd legacy plus pump, the pump exhibited "no disposable, pump won't run" alarm.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection showed the device was in good condition.A review of the event history log (ehl) identified no disposable alarm.During the functional testing the reported issue was unable to be duplicated.The root cause of the issue was unable to be determined.As a preventative measure, the downstream sensor was replaced, and the upstream sensor re-calibrated.
|
| |
|