| Catalog Number |
814311400 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2021-03170.Concomitant medical products: item# 211201202; lot# unk.Item# 211201200; lot# unk.Report source: foreign - event occurred in (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that surgeon completed a hip fracture nailing operation using the affixus hfn.At the end of the procedure, the jig connecting bolt was unable to be removed from the nail.They ended up using two trays and the handle of insertion jig instrument broke.In the end they had to remove the whole prosthesis and open a third set and new nail to complete the surgery.There was a delay in the procedure of an unknown time.Attempts have been made and no further information has been provided.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; g3; g6; h1; h2; h3; h6 visual examination of the provided pictures identified the nail is still connected to the targeting jig.The insertion jig bolt is not visible in the provided pictures.Part and lot information cannot be confirmed.Medical records were not provided.Review of the device history record(s) for part 814311400 identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|