| Model Number |
MR8-F2/7TA23 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No conclusion can be drawn.No evaluation was performed, as the tool was discarded.No evaluation was performed for the em800 and mr8-as09, as these devices were not returned.If these devices were returned in the future, product analysis may be performed.The user manual contains the following warning ¿do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff.¿ we will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that while using an em800 and mr8-as09 for a craniotomy procedure, the tool was broken and a small part of the broken tool remain inside the motor.It was reported that the biggest part of the broken tool was removed from the patient.There was also a 20-minute procedure delay and another em800 was used to continue the craniotomy and finish the surgery.No patient impact reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The em800 has been received and analyzed.Evaluation determined that there was a part of the tool stuck inside the motor.The likely cause of failure could not be determined.There is no evidence that any other failures during analysis were found that may have contributed to this event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
On follow-up, it was confirmed that there was no patient or staff impact.It was also reported that the patient had no further issues post-surgery.
|
| |
|