• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA NEUROSCICENCS IMPLANTS SA OSVII, CONNECTOR WITH ANTECHAMBER.; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA NEUROSCICENCS IMPLANTS SA OSVII, CONNECTOR WITH ANTECHAMBER.; N/A Back to Search Results
Model Number 909712
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 10/26/2021
Type of Reportable Event Serious Injury
Event or Problem Description
A facility reported that on (b)(6) 2021, an osvii valve (909712) was implanted for hydrocephalus (post tumor resection).After 3 days of implantation, the patient's hydrocephalus was not improved and a new emergency surgical intervention was required on (b)(6) 2021 to check the implanted valve.In the revision surgery, it was verified that the osvii valve was not draining the cerebrospinal fluid (csf).The medical team chose to replace the osvii valve with the codman hakim medium pressure and the csf drainage worked correctly.The surgery was performed following the manufacturer's guidelines.Patient had neurological worsening, increased ventricular ectasia, underwent reoperation, and as a result, her hospital stay increased by 7 days.Patient is showing improvement with the implantation of the new valve.
 
Additional Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10 osvii,connector antechamber (909712) was returned for evaluation: dhr review - the device history records of ref 909712, lot 217615, sn (b)(6), were reviewed and did not reveal any anomaly that could explain the reported event.Failure analysis - visual inspection of the valve under magnification revealed no anomaly.The valve was pressure/flow tested and found functional, but slightly over specification complaint reported the valve was not draining, however, the received valve investigation does not verify the complaint, valve is not obstructed and does not underdrain.Root cause analysis - the exact cause of the reported underdrainage of the valve could not be determined by the investigation.The received valve investigation does not verify the complaint, valve is not obstructed and does not underdrain.Ventricular catheter was not returned and may have been obstructed.Valve manipulations after explantation may explain the slight shift of the pressure/flow curve.Given the device analysis, no further investigation nor corrective action is deemed required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSVII, CONNECTOR WITH ANTECHAMBER.
Common Device Name
N/A
Manufacturer (Section D)
INTEGRA NEUROSCICENCS IMPLANTS SA
2905 route des dolines
2905 route des dolines
sophia antipolis F-069 21
FR  F-06921
Manufacturer (Section G)
INTEGRA NEUROSCICENCS IMPLANTS SA
2905 route des dolines
sophia antipolis F-069 21
FR   F-06921
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12845043
Report Number9612007-2021-00037
Device Sequence Number17216529
Product Code JXG
UDI-Device Identifier10381780034032
UDI-Public(01)10381780034032
Combination Product (Y/N)N
PMA/510(K) Number
K140722
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/06/2023
Device Model Number909712
Device Catalogue Number909712
Device Lot Number0217615
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/28/2021
Supplement Date Received by Manufacturer03/11/2022
Initial Report FDA Received Date11/19/2021
Supplement Report FDA Received Date03/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-