Model Number HST III SEAL (4.5MM), 5-PACK |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.Device discarded.
|
|
Event Description
|
The hospital reported that during a coronary artery bypass procedure using hst iii seal (4.5mm), 5-pack, the seal could not be loaded, and the procedure was completed without any problems using a spare heart string.No health hazard to the patient.
|
|
Manufacturer Narrative
|
Trackwise # (b)(4).Corrected section: h3-if other provide code -explain- device not returned.Analysis of production: (3331/213/67) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure mode.Historical data analysis for similar complaints: (4109/213/67)the review of the historical data indicates that no other similar complaints was reported for the same lot number and reported failure mode.Trend analysis: (4110/213/67) the overall 24 month product complaint trend data for the period dec-2019 through nov-2021 was reviewed.There were no triggers identified for the review period.
|
|