| Model Number |
NC055-040-4 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Malunion of Bone (4529)
|
| Date of Event |
10/29/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The root cause appears to have been that two screws experienced excessive load either by traumatic fracture or improper implanting.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon who initially implanted hardware scheduled a revision of centerline.Surgeon's colleague assisted and decided they wanted to remove all hardware and replace with a different product upon exposure and removing hardware, they noticed one complete side of screws (3) had broken.Other side intact.Hardware is being washed, collected & returned.
|
| |
|