MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS PHOENIX; CATHETER, RECORDING, ELECTRODE, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS PHOENIX; CATHETER, RECORDING, ELECTRODE, REPROCESSED Back to Search Results
Model Number 401904
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device, a bend was identified 3 cm from the distal tip; which is distal to the transition point.The handle, steering knob, and connector pins appear to be intact.The device was evaluated.The device did not meet the curve or planarity specification.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to a bend distal to the transition point as a result of mishandling subsequent to distribution, including shipping/storage conditions or improper manipulation of the catheter.The instructions for use (ifu) state: inspect the catheter for overall condition and physical integrity.Do not use the catheter if electrodes appear loose or if any damage is noted.If such problems exist, return the catheter and packaging to stryker sustainability solutions.Connect the catheter to the corresponding cable connector.Connect the cable to the correct electronic equipment for recording and/or sensing.Observe polarity of proximally located connector pins of the interface cable when connecting to the electronic equipment.Isolate any unused connector pins to reduce development of accidental current pathways to the heart.Follow a suitable electrophysiology study protocol.Do not autoclave catheter.Do not use for electrical ablation.Avoid excessive contact of handpiece with fluids, as this could adversely affect the electrical performance of the catheter.Diagnostic ep catheters are not recommended for long term pacing.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported catheter would not deflect.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
CATHETER, RECORDING, ELECTRODE, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS PHOENIX
10232 s. 51st st.
phoenix AZ 85044
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS PHOENIX
10232 s. 51st st.
phoenix AZ 85044
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key12846794
Report Number0002090040-2021-00042
Device Sequence Number19921185
Product Code NLH
UDI-Device Identifier00885825007263
UDI-Public00885825007263
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K052603
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/19/2023
Device Model Number401904
Device Catalogue Number401904RR
Device Lot Number4404644
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/20/2021
Initial Report FDA Received Date11/19/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-