| Model Number |
0998-00-0800-53 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information is being requested with regard to the repair and status of the iabp.A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during use on a patient, the cs100 intra-aortic balloon pump (iabp) gave a gas loss alarm.The unit was swapped out with another to continue therapy.No patient harm, serious injury or adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Testing of actual/suspected device (10/213): a getinge field service engineer (fse) was dispatched to evaluate the iabp.To fix the issue, the fse started unit and ran on a test balloon no indication of helium leaking.The unit went through the helium leak tests as per the service manual, all passed and within factory specs.The fse reassembled the unit and it ran on the test balloon with no issues and a complete calibration and electrical safety test.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) gave a gas loss alarm.The unit was swapped out with another to continue therapy.No patient harm, serious injury or adverse event was reported.
|
| |
|
Event or Problem Description
|
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) gave a gas loss alarm.The unit was swapped out with another to continue therapy.No patient harm, serious injury or adverse event was reported.
|
| |
|