MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED2-500-35
Device Problems Unintended Ejection (1234); Activation Failure (3270); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/08/2021
Event Type  malfunction  
Manufacturer Narrative
A separate report will be submitted regarding the ped2-500-30.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a pipeline flex with shield that failed to open and then because stuck in the phenom microcatheter and a second pipeline flex with shield that also failed to open.The patient was undergoing a procedure to treat a ruptured saccular left internal carotid artery (ica) supra clinoid aneurysm.The aneurysm max diameter was 5.91mm and the neck diameter was 4.23mm.The distal landing zone was 3.6mm and the neck diameter was 4.75mm.Vessel tortuosity was reported to be normal.It was reported that it was planned to deploy a pipeline from the middle cerebral artery (mca) to the ica.The pipeline flex (ped2-500-35) was delivered and deployment started but the device did not open at the distal end.The device was resheathed and redeployed but still did not open distally so it was decided to removed the pipeline.However it deployed and became stuck in the phenom 27 microcatheter.Another pipeline flex (model: ped2-500-30) was then tried with a new phenom 27 microcatheter.However once delivered, the ped2-500-30 also failed to open at the distal end.The device was removed.A pipeline (ped2-500-25) was then used which deployed and was implanted successfully.There was no harm or injury to the patient.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received reported the pushwire was not rotated or pulled back.The pipeline was not in a vessel bend when the failure to open occurred; it was in a straight segment of the middle cerebral artery (mca).Resheathing was done to try and open the pipeline without success; no other devices or maneuvers were tried to open the pipeline.
 
Manufacturer Narrative
H3.Product analysis: equipment used: video inspection system (m-85519), ruler (m-83361), camera (panasonic lumix dmc-zs5), in-house 0.0265in mandrel.As found condition: the pipeline flex shield was returned inside of the phenom 27 catheter; within the inner pouch and outer carton; inside of a sealed bio-hazard bag and a shipping box.Visual inspection/damage location details: the distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.The distal hypotube appeared to be stretched with the ptfe shrink tubing still intact.The distal end of the braid was not opened due to damaged braid.The proximal end of the braid were found fully opened and frayed.Bends were found at 20.0cm to 28.5cm from the proximal end.No defects were found with the tip coil, distal marker, re-sheathing marker, re-sheathing pad or with the proximal bumper.The total and usable lengths of the catheter were measured to be within specifications.The catheter tip and marker were examined; no damages were found.The catheter body appeared to be accordioned at 8.0cm to 15.5cm from the distal tip.No flash or voids molded were observed in the hub.No other anomalies were observed.Testing/analysis: the pipeline flex shield could not be pushed forward or removed.For further examination, the catheter was cut to remove the pipeline flex shield delivery system.The catheter was flushed with water and found patent.The catheter was then tested by running an in-house 0.0265¿ mandrel through catheter tip and hub.The mandrel successfully passed through the catheter hub with no issues; however, resistance was observed at the damaged locations.Conclusion: based on the returned devices, the customer's complaint was confirmed as the pipeline flex shield was found stuck inside the phenom catheter.The distal end of the braid was not opened due to damaged braid.The damage to the braid on the ends of the pipeline flex shield is likely the results of the physician re-sheathing the device more than recommended two times.From the damages seen on the catheter (accordioning), pushwire (bending), pipeline flex shield braid (fraying) and hypotube (stretching); it appears there was high force used.It is likely these damages occurred when the customer attempted to advance/retrieve the pipeline flex shield through the phenom catheter against the resistance.However, the cause for resistance could not be determined.Possible causes include patient's vessel tortuosity and lack of continuous flush with heparinized saline during procedure.There was no non-conformance to specifications identified that led to the reported issues.Per our instructions for use (ifu), the user should: ¿discontin ue delivery of the device if high force or excessive friction is encountered during delivery.Identify the cause of the resistance and remove device and microcatheter simultaneously.Advancement of the ped against resistance may result in device damage or patient injury.Never advance or withdraw an intralumenal device against resistance until the cause of resistance is determined by fluoroscopy.If the cause cannot be determined, withdraw the catheter.Movement of the micro catheter against resistance may result in damage to the micro catheter, or the vessel.Do not use inpatients in whom the angiography demonstrates the anatomy is not appropriate for endovascular treatment, due to conditions such as severe intracranial vessel tortuosity or stenosis.¿.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX W/SHIELD TECHNOLOGY
Type of Device
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key12848633
MDR Text Key285549895
Report Number2029214-2021-01506
Device Sequence Number1
Product Code OUT
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P100018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPED2-500-35
Device Catalogue NumberPED2-500-35
Device Lot NumberB213508
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/28/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
Patient Weight73 KG
-
-