Model Number URF-V3R |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device to omsc was returned to omsi for evaluation.(b)(4) checked the subject device and found the reported phenomenon.The exact cause has been under investigation.Therefore, the exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair of the subject device at olympus (b)(4), it was found that the bending section cover of the subject device had been torn and the broken internal metal part had been sticking out from the inside of bending section cover of the subject device.The occurrence date of the event is unknown, and there was no report of patient injury associated with this event.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.The subject device was returned to olympus medical systems india (omsi).Olympus medical systems corp.(omsc) reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the information from omsi and similar past cases, omsc surmised that this phenomenon was attributed to the excessive force being applied to the bending section due to the user handling.If additional information is received, this report will be supplemented.
|
|