MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V3R
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device to omsc was returned to omsi for evaluation.(b)(4) checked the subject device and found the reported phenomenon.The exact cause has been under investigation.Therefore, the exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair of the subject device at olympus (b)(4), it was found that the bending section cover of the subject device had been torn and the broken internal metal part had been sticking out from the inside of bending section cover of the subject device.The occurrence date of the event is unknown, and there was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device was returned to olympus medical systems india (omsi).Olympus medical systems corp.(omsc) reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the information from omsi and similar past cases, omsc surmised that this phenomenon was attributed to the excessive force being applied to the bending section due to the user handling.If additional information is received, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12850862
MDR Text Key283255568
Report Number8010047-2021-14877
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberURF-V3R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/26/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-