| Model Number |
2544-00-026 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # = > (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the instrument to be worn.The part that guides the saw is worn.An exact cutting-preparation is no longer possible.No adverse patient consequences, no surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: the examination of the returned device can confirm the reported allegation.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: 254400026 / abb97158.- quantity manufactured: 234.- date of manufacture: 17 dec 2013.- any anomalies or deviations identified in dhr: none.- expire date: n/a nonsterile reusable.- instrument ifu reference: (b)(4).Device history review: 254400026 / abb97158.- quantity manufactured: 234.- date of manufacture: 17 dec 2013.- any anomalies or deviations identified in dhr: none.- expire date: n/a nonsterile reusable.- instrument ifu reference: (b)(4).
|
| |
|