APOLLO ENDOSURGERY, INC. X-TACK ENDOSCOPIC HELIX TACKING SYSTEM; ENDOSCOPIC TISSUE APPROXIMATION DEVICE Back to Search Results
Model Number XTACK-235-H
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
Attempted to close defect from emr (endoscopic mucosal resection) with apollo x-tack suture system.Device would not deploy properly.Provider was able to implant one helix, but unable to remove the catheter without pulling helix out of tissue.Called vendor rep for assistance, and watched videos of device.Still unable to get device to work properly.Finished procedure without closing defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-TACK ENDOSCOPIC HELIX TACKING SYSTEM
Common Device Name
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s capital of tx hwy
bldg 1, ste 300
austin TX 78746
MDR Report Key12851324
Report Number12851324
Device Sequence Number17145215
Product Code OCW
UDI-Device Identifier10811955020770
UDI-Public(01)10811955020770(17)211205(10)AF04218
Combination Product (Y/N)N
Initial Reporter StateFL
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 11/18/2021
Report Date (Section F) 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberXTACK-235-H
Device Catalogue NumberXTACK-235-H
Device Lot NumberAF04218
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age0 YR
Event Location Hospital
Date Report to Manufacturer11/22/2021
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date11/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexFemale
Patient Weight73 KG
Date Report Sent to FDA11/18/2021
-
-