• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERYAN MEDICAL LTD BIOMIMICS 3D VASCULAR STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERYAN MEDICAL LTD BIOMIMICS 3D VASCULAR STENT SYSTEM Back to Search Results
Model Number 142122-09
Medical Device Problem Codes Fracture (1260); Material Deformation (2976)
Health Effect - Clinical Codes Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 11/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complaint investigation is ongoing.The device was not returned for evaluation.Angiography imaging will be reviewed as part of the investigation.If further information regarding this event becomes available, or the investigation is concluded, a followup report will be submitted.
 
Event or Problem Description
On (b)(6) 2021, a physician reported to a veryan representative information about a stent fracture identified in a patient who was implanted in (b)(6) 2021.The fracture was identified during a follow-up angiogram on (b)(6) 2021.The veryan representative received additional information on 02 november 2021 stating that angioplasty and placement of another stent inside the biomimics was performed.The stent remains implanted.Angiographic imaging is available.It is not known if the physician believes the incident is related to the device.
 
Additional Manufacturer Narrative
A detailed review of all the lot history records pertaining to the relevant stent and delivery system lots showed no issues that were deemed related to the complaint investigation.The investigation included communication/interviews and review of smartphone images provided as well as additional information from the reporting physician.The images were not the original angiographic images.The originals were requested but the clinical site clarified that the original images would not be made available.Based on a review of the images that were available, it was possible to detect a distorted stent pattern in the biomimics 3d stent.Determination of a stent fracture and the extent of stent pattern distortion was not possible from the images provided.Additional information from the physician highlighted that the biomimics stent was implanted using a contralateral approach, balloon angioplasty was performed prior to stent placement.There were two stents implanted at the index procedure, the biomimics 3d stent was placed distally to the viabahn stent.The physician also clarified that the procedure was not out of the ordinary and there were no issues to report.The complaint was categorised as a distorted stent pattern and a root cause could not be determined.The complaint is not related to a deficiency of the device.Any additional information provided will be provided in a follow-up report.Section h.6.Adverse event coding has been updated to reflect the findings and conclusion of the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Common Device Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Manufacturer (Section D)
VERYAN MEDICAL LTD
unit 11 galway technology park
parkmore road
galway city, H91 V E0H
EI  H91 VE0H
Manufacturer Contact
alan mcdonagh
unit 11 galway technology park
parkmore road
galway, H91 V-E0H
EI   H91 VE0H
MDR Report Key12851921
Report Number3011632150-2021-00045
Device Sequence Number17145088
Product Code NIP
UDI-Device Identifier05391526850442
UDI-Public(01)05391526850442(17)220314(11)200813(10)48415
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P180003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/14/2022
Device Model Number142122-09
Device Catalogue Number142122-09
Device Lot Number48415
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2021
Supplement Date Received by Manufacturer12/21/2021
Initial Report FDA Received Date11/22/2021
Supplement Report FDA Received Date01/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-