WILLIAM COOK EUROPE ZENITH TX2 DISSECTION ENDOVASCULAR GRAFT STRAIGHT COMPONENT; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number G47442
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Description of event according to the initial reporter: post deployment of zdeg, green sleeve and trigger wires would not release.Excessive force required to release trigger wires causing dilator and device to migrate distally resulting in encroachment of the celiac artery.Subsequent stent implantation was required to preserve celiac artery.Patient outcome: no unintended device remained inside the patient´s body.The patient did require an additional procedure due to occurrence.A celiac covered stent was placed to protect celiac from stent migration and mal-perfusion.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZENITH TX2 DISSECTION ENDOVASCULAR GRAFT STRAIGHT COMPONENT
Common Device Name
MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key12852561
Report Number3005580113-2021-00142
Device Sequence Number19793382
Product Code MIH
UDI-Device Identifier10827002474427
UDI-Public(01)10827002474427(17)240302(10)E4079630
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 11/22/2021
Report Date (Section F) 11/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/02/2024
Device Model NumberG47442
Device Catalogue NumberZDEG-P-28-142-PF-US
Device Lot NumberE4079630
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Event Location Hospital
Date Report to Manufacturer11/19/2021
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date11/22/2021
Date Device Manufactured03/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight73 KG
Date Report Sent to FDA11/22/2021
-
-