MAUDE Adverse Event Report: ARTHREX, INC. IMPL SYS,CMC MINI T-ROPE,1.1 MM; WASHER, BOLT NUT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. IMPL SYS,CMC MINI T-ROPE,1.1 MM; WASHER, BOLT NUT Back to Search Results
Model Number IMPL SYS,CMC MINI T-ROPE,1.1 MM
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during a hand trapezectomy surgery the eyes of the 2 pins of the device broke inside the patient.All broken parts were retrieved from the patient.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a different device.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Complaint confirmed.Visual evaluation showed one of the k-wires had the tip and shaft bent.The nylon wire is broken.Other k-wire had the proximity damaged and nylon wire missing.Critical dimensions or functioning could not be performed due to the damage of the devices.A likely cause of the deformation is user-applied mechanical forces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL SYS,CMC MINI T-ROPE,1.1 MM
Common Device Name
WASHER, BOLT NUT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key12853169
Report Number1220246-2021-03966
Device Sequence Number4394712
Product Code HTN
UDI-Device Identifier00888867048966
UDI-Public00888867048966
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K140328
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model NumberIMPL SYS,CMC MINI T-ROPE,1.1 MM
Device Catalogue NumberAR-8919DS
Device Lot Number13854200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2021
Supplement Date Received by Manufacturer11/02/2021
Initial Report FDA Received Date11/22/2021
Supplement Report FDA Received Date01/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-