• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. APOLLO RF 50°; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. APOLLO RF 50°; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number APOLLO RF 50°
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During returned device evaluation, a reportable malfunction was discovered.The complaint is confirmed.Device was received: ar-9815 batch 2009082, unpackaged.Functional testing was not performed due to the device's discovered reportable status.The observed burn marks at the distal tip suggest excessive ablation without proper fluid content, making it the most likely cause of the reported issue.
 
Event or Problem Description
It was reported that during a knee arthroscopy the device stopped working.No part of the device broke off.There was no harm or adverse event for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.During returned device evaluation, a reportable malfunction was discovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APOLLO RF 50°
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key12853664
Report Number1220246-2021-03982
Device Sequence Number20289497
Product Code GEI
UDI-Device Identifier00888867142527
UDI-Public00888867142527
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K161581
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model NumberAPOLLO RF 50°
Device Catalogue NumberAR-9815
Device Lot Number2009082
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2021
Initial Report FDA Received Date11/22/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-