| Model Number |
29MECJ-502 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2021, a 29mm sjm masters series valve expanded cuff was implanted.Post-procedure, after the patient was closed, a large amount of blood was drained and ventricular rupture was observed.As a result, the masters valve was explanted.The user alleged that injury to the myocardium by a non-abbott device was the cause of the ventricular rupture.A new 27mm sjm masters series valve expanded cuff was successfully implanted.The patient was hemodynamically stable throughout the procedure and no clinically significant prolonged procedure time was reported.The patent was reported to be in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of ventricular rupture was reported.The investigation confirmed the device met visual specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined, however information from the field indicated that the user alleged the injury was due to a non-abbott device.
|
| |
|