• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE STRYKER 2-0 KNOT PUSHER/SUTURE CUTTER AND SLOTTED CANNULA SET; ACCESSORIES, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE STRYKER 2-0 KNOT PUSHER/SUTURE CUTTER AND SLOTTED CANNULA SET; ACCESSORIES, ARTHROSCOPIC Back to Search Results
Model Number 4721
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the planned meniscus repair was aborted and the surgeon opted for a meniscectomy.
 
Event or Problem Description
It was reported that the planned meniscus repair was aborted and the surgeon opted for a meniscectomy.
 
Additional Manufacturer Narrative
The device manufacturer date is not known.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: premature cutting of suture.Probable root cause: design: device opening for suture too large (knot slips through).Inadequate material selection for cutter (distal or proximal features).Distal assembly of cutter susceptible to fragmentation.Inadequate handle spring force.Process: inadequate assembly of device.Inadequate deburring/edge break on non-cutting surfaces.Device not manufactured to specification.Application: lack of due care when inserting device into soft tissue.Excessive force.The reported failure mode will be monitored for future re-occurrence.H3 other text: 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRYKER 2-0 KNOT PUSHER/SUTURE CUTTER AND SLOTTED CANNULA SET
Common Device Name
ACCESSORIES, ARTHROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key12857922
Report Number0002936485-2021-00661
Device Sequence Number4080620
Product Code NBH
UDI-Device Identifier07613327364699
UDI-Public07613327364699
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4721
Device Catalogue Number4721
Device Lot Number006442
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/28/2021
Supplement Date Received by Manufacturer10/28/2021
Initial Report FDA Received Date11/22/2021
Supplement Report FDA Received Date01/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-