MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE 1688 CAMERA CONTROL UNIT (CCU) - INTERNATIONAL KIT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE 1688 CAMERA CONTROL UNIT (CCU) - INTERNATIONAL KIT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 1688010000
Device Problem Poor Quality Image (1408)
Patient Problems Discomfort (2330); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/31/2021
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was conversion to open procedure.
 
Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: image flipped.Probable root cause: cables; hdmi cables; connectors; digital board; power supply; filter / fuse; ac inlet board; main board; transition board; intermittence between transition board and ch connector; software; camera head; coupler; problem toggling light source; connected monitors; leaking; poor grounding; no/incorrect communication with light source; soaking cap disconnection during sterilization; electromagnetic interference from rf communication, hf surgical instruments, esd, or power surge; cybersecurity attack; pendulum ch inertial measurement unit ; incident light from surgical scene is reflected by coupler/ch window; use error.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
 
Event Description
It was reported that there was conversion to open procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1688 CAMERA CONTROL UNIT (CCU) - INTERNATIONAL KIT
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
hannah stalnaker
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key12858207
MDR Text Key281132392
Report Number0002936485-2021-00673
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier07613327420081
UDI-Public07613327420081
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K182160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1688010000
Device Catalogue Number1688010000I
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/31/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-