NEOTRACT, INC. NEOTRACT UROLIFT SSYTEM; UROLIFT SYSTEM Back to Search Results
Model Number UROLIFT SYSTEM
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2021
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2021, neotract was made aware of a patient who received a successful prostatic urethral lift (pul) procedure on (b)(6) 2021.During the procedure, it was noted that one device did not properly deploy causing the capsular tab to break.On (b)(6) 2021, returned device analysis confirmed that a 0.25 inch fragment was missing from the capsular tab.The physician was informed and stated that no additional scanning will be performed and the fragment was likely flushed out during the procedure.No patient harm or injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOTRACT UROLIFT SSYTEM
Common Device Name
UROLIFT SYSTEM
Manufacturer (Section D)
NEOTRACT, INC.
4155 hopyard road
pleasanton CA 94588
Manufacturer (Section G)
NEOTRACT, INC.
4155 hopyard road
pleasanton CA 94588
Manufacturer Contact
brian gall
4155 hopyard rd
pleasanton, CA 94588
9253296547
MDR Report Key12858433
Report Number3015181082-2021-00039
Device Sequence Number19912749
Product Code PEW
UDI-Device Identifier10814932020152
UDI-Public10814932020152
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K193269
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUROLIFT SYSTEM
Device Catalogue NumberIPN056625
Device Lot NumberP00510
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2021
Initial Report FDA Received Date11/22/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-