• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RXSIGHT, INC. LIGHT ADJUSTABLE LENS (LAL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RXSIGHT, INC. LIGHT ADJUSTABLE LENS (LAL) Back to Search Results
Model Number 60005
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code Visual Impairment (2138)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Site reported that the light adjustable lens was explanted as the desired refractive outcome was not achieved after light adjustment treatments.Rxsight's awareness of event was (b)(6) 2021.Device history record for the lens was reviewed.No issues were noted.The explanted lens was returned for evaluation.Visual and optical testing verified the presence of a zone near the center of the lens.The presence of the zone is indicative of lens exposure to ambient uv light.
 
Event or Problem Description
Site reported that the light adjustable lens was explanted as the desired refractive outcome was not achieved after light adjustment treatments.Rxsight's awareness of event was (b)(6) 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT ADJUSTABLE LENS (LAL)
Common Device Name
LIGHT ADJUSTABLE LENS (LAL)
Manufacturer (Section D)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer (Section G)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer Contact
kelli carrillo
100 columbia
aliso viejo, CA 92656
MDR Report Key12860944
Report Number3012712027-2021-00077
Device Sequence Number19899852
Product Code PZK
UDI-Device Identifier00818806020159
UDI-Public010081880602015917220930
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P160055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model Number60005
Device Catalogue Number60005
Device Lot NumberL02-001336
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/26/2021
Initial Report FDA Received Date11/23/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/04/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-