| Model Number |
60005 |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Code |
Visual Impairment (2138)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Site reported that the light adjustable lens was explanted as the desired refractive outcome was not achieved after light adjustment treatments.Rxsight's awareness of event was (b)(6) 2021.Device history record for the lens was reviewed.No issues were noted.The explanted lens was returned for evaluation.Visual and optical testing verified the presence of a zone near the center of the lens.The presence of the zone is indicative of lens exposure to ambient uv light.
|
| |
|
Event or Problem Description
|
|
Site reported that the light adjustable lens was explanted as the desired refractive outcome was not achieved after light adjustment treatments.Rxsight's awareness of event was (b)(6) 2021.
|
| |
|