| Model Number |
NSEAL525RH |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: v9481m.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the nseal525rh device was received with the top of the i-blade fractured and slightly lifted.The device was connected to the generator and it was recognized.Because the i blade was damaged not all functional testing could be performed with the generator.The condition of the blade prevented the functionality of the jaw.The jaw was unable to cycle open and close.A probable cause of the damage to the i blade could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.As described in the ifu: do not use excessive force on the closing handle to close the jaws; grasp only as much tissue as will fit between the jaws where the current will pass.Greater amounts of tissue require more closing handle force.Excessive force could damage the device.Please reference the instruction for use for more information: as part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
|
| |
|
Event or Problem Description
|
|
It was reported that during the anal exam under anesthesia the enseal device jaws stopped opening and closing.Case completed with another like device.There were no patient consequences reported.
|
| |
|
Search Alerts/Recalls
|