Upon receipt, the pacemaker was interrogated and the memory content was analyzed.The battery status showed 40 percent.During the analysis of the available iegms noise was observed in the right ventricular channel and the inspection revealed a decreasing ventricular lead impedance.The header of the device was analyzed.The set screws could be easily screwed in and out, there was no foreign material inside the header bores.All dimensions of the header bores were within the range requested by the standard specifications.Also, the spring elements of the pacemaker did not show any deviations.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.There was no indication of a device malfunction.An additional long-term pacing test was initiated.During the test, each pacing pulse was recorded.The evaluation of these pacing pulses documented regular device behavior.There was no intermittent or permanent loss of output.In conclusion, the device is fully functional.The analysis did not show any deviations from the technical specifications.The analysis did not reveal any sign of a material or manufacturing problem.
|