• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ONKOS SURGICAL ELEOS; CEMENTED STEM, 15X120MM, STRAIGHT, FLUTED, 32MM COLLAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL ELEOS; CEMENTED STEM, 15X120MM, STRAIGHT, FLUTED, 32MM COLLAR Back to Search Results
Model Number CS-15120-03M
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Post Operative Wound Infection (2446)
Date of Event 10/31/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient developed an alleged wound infection and underwent a revision surgery on (b)(6) 2021.During the revision surgery, the following eleos implants were revised: tibial hinge component, distal femur axial pin, and tibial poly spacer.The following eleos implants remain implanted from the patient's original surgery: distal femur, male-female midsection, cemented stem extension, tibial baseplate, and cemented segmental stem.The surgeon performed a washout.No additional information regarding this adverse event has been provided.
 
Additional Manufacturer Narrative
The investigation is in process.Once the investigation is complete, a supplemental mdr will be submitted accordingly.Multiple mdrs have been submitted for this adverse event: #3013450937-2021-00301, #3013450937-2021-00302, #3013450937-2021-00303, #3013450937-2021-00304, #3013450937-2021-00305, #3013450937-2021-00306, #3013450937-2021-00307.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.H3: device evaluated by manufacturer updated to no; h6: type of investigation code updated to 3331: analysis of production records; h6: type of investigation code updated to 4111: communication/interviews; h6: type of investigation code updated to 4110: trend analysis; h6: type of investigation code updated to 4114: device not returned; h6: type of investigation code updated to 4117: device not accessible for testing; h6: type of investigation code updated to 4109: historical data analysis; h6: investigation findings code updated to 213: no device problem found; h6: investigation conclusions code updated to 4315: cause not established; h10: additional narratives/data.
 
Event or Problem Description
It was reported that the patient developed an alleged wound infection and underwent a revision surgery on (b)(6) 2021.During the revision surgery, the following eleos implants were revised: tibial hinge component, distal femur axial pin, and tibial poly spacer.The following eleos implants remain implanted from the patient's original surgery: distal femur, male-female midsection, cemented stem extension, tibial baseplate, and cemented segmental stem.The surgeon performed a washout.No additional information regarding this adverse event has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
CEMENTED STEM, 15X120MM, STRAIGHT, FLUTED, 32MM COLLAR
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer (Section G)
PHILLIPS PRECISION INC
7 paul kohner pl
elmwood park NJ 07407
Manufacturer Contact
sara dailey
77 east halsey road
parsippany, NJ 07054
MDR Report Key12865358
Report Number3013450937-2021-00308
Device Sequence Number19192787
Product Code KRO
UDI-Device IdentifierB278CS1512003M0
UDI-PublicB278CS1512003M0
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K161520
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberCS-15120-03M
Device Catalogue NumberCS-15120-03M
Device Lot Number85223
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2021
Supplement Date Received by Manufacturer01/27/2022
Initial Report FDA Received Date11/23/2021
Supplement Report FDA Received Date01/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/21/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P/N 25000009E, ELEOS DISTAL FEMUR.; P/N 25001040E, ELEOS MALE-FEMALE MIDSECTION.; P/N 25001212E, ELEOS TIBIAL POLY SPACER.; P/N 25002111E, ELEOS DISTAL FEMUR AXIAL PIN.; P/N KSC01265E, ELEOS CEMENTED STEM EXTENSION.; P/N TB-2202E-01M, ELEOS TIBIAL BASEPLATE.; P/N THSMWRS01M, ELEOS TIBIAL HINGE COMPONENT.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age89 YR
Patient SexFemale
-
-