According to the reporter, during a laparoscopic thoracoscopic lobectomy of the lung, after attaching the cartridge to the adapter using the power handle and removing the yellow tab from the cartridge in order to check the opening and closing of the jaws, the user noticed that the protrusion part of the yellow tab was damaged and broke off.It was noted no damage site was found.It was confirmed that the foreign object had not fallen into the surgical field, and the operation was continued as usual.There was no patient injury.
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the shipping wedge was damaged.It was reported that the shipping wedge was broken, or yellow pieces have broken off from the shipping wedge.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if the shipping wedge pull tab is pulled in a lateral fashion towards the distal end of the single use loading unit (sulu) channel rather than away from the channel, or if the loading unit is clamped prior to removal of the yellow shipping wedge.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: instructions for proper loading or reload and it clearly states instructions for proper use of stapler.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|