| Model Number |
C8303 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event unit returned to applied medical for evaluation.A follow-up report will be provided upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Name of procedure being performed na (incoming inspection) detailed description of event: [name] incoming inspection po#(b)(4).We would like to inform you that some of the products didn't pass our receiving inspection.There is a item require cer in this report.Particulate: hair, bug - c8303 lot# 1424309 (1 ea).Additional information received via email 01nov2021 from [name] regulatory affairs and quality assurance: it was reported that there was only one bug found.No report of hair in packaging.Patient status: na (no patient involvement).Type of intervention: na (incoming inspection).
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the presence of a bug inside the sterile unopened pouch.Based on the condition of the returned unit and the description of the event, the bug was most likely located inside a clear film roll used for packaging and was not detected during the manufacturing process.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
|
| |
|
Event or Problem Description
|
|
Name of procedure being performed.Na (incoming inspection).Detailed description of event: [name] incoming inspection (b)(4).We would like to inform you that some of the products didn't pass our receiving inspection.Please find attached spreadsheet.There is a item require cer in this report.Particulate: hair, bug - c8303 lot# 1424309 (1 ea).Additional information received via email 01nov2021 from [name] regulatory affairs and quality assurance: it was reported that there was only one bug found.No report of hair in packaging.Patient status: na (no patient involvement).Type of intervention: na (incoming inspection).
|
| |
|