• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/05/2021
Event Type  malfunction  
Event Description
It was reported that the the cs300 intra-aortic balloon pump (iabp) had an error code 5 battery issue.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.
 
Manufacturer Narrative
Additional information is being requested with regard to the repair and status of the iabp.A supplemental report will be submitted if this information is provided to us.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, h2, h10, h11 corrected fields: d10, e1 (event site email), h3, h6 (type of investigation, component codes) testing of actual/suspected device (10/3233): a getinge field service engineer (fse) was dispatched to investigate.The fse turned on the iabp unit and received a maintenance code #5 message on the main screen.The fse reviewed the iabp log files and observed many error codes as well as some unknown error codes.The fse calibrated the helium transducer and replaced the datasettes.Subsequently, the fse performed a system check and calibration verification per the service manual, then ran the iabp unit on a trainer.The iabp unit passed all testing and was cleared for use and returned to the customer.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
N/a.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12866481
MDR Text Key284447783
Report Number2249723-2021-02736
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 02/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/03/2014
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-