Updated fields: b4, g4, g7, h2, h10, h11 corrected fields: d10, e1 (event site email), h3, h6 (type of investigation, component codes) testing of actual/suspected device (10/3233): a getinge field service engineer (fse) was dispatched to investigate.The fse turned on the iabp unit and received a maintenance code #5 message on the main screen.The fse reviewed the iabp log files and observed many error codes as well as some unknown error codes.The fse calibrated the helium transducer and replaced the datasettes.Subsequently, the fse performed a system check and calibration verification per the service manual, then ran the iabp unit on a trainer.The iabp unit passed all testing and was cleared for use and returned to the customer.A supplemental report will be submitted upon completion of our investigation.
|