FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW + TRACHEOSTOMY INTERFACE; BTT Back to Search Results
Model Number OPT970
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A distributor in (b)(6) reported via a fisher & paykel healthcare (f&p) field representative, that the tubing of an (b)(4) optiflow + tracheostomy direct connection interface was found broken before use.There was no patient involvement.
 
Additional Manufacturer Narrative
(b)(4).The complaint (b)(4) optiflow + tracheostomy direct connection interface is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of our investigation.
 
Additional Manufacturer Narrative
(b)(4).The opt970 optiflow + tracheostomy direct connection interface is used to deliver humidified oxygen to patients via tracheostomy.The interface is held in place by a neck strap and also includes a lanyard which is placed around the patient's neck or attached to the patient's clothing or bedding to remove the load of the breathing circuit from the patient's trache.Method: the complaint opt970 optiflow + tracheostomy direct connection interface was returned to fisher & paykel healthcare (f&p) new zealand for investigation, where it was visually inspected.Results: visual inspection of the returned cannula revealed that the tubing was stretched and pulled apart.Conclusion: we are unable to determine the cause of the reported damage to the subject opt970 optiflow + tracheostomy direct connection interface.However, the damage observed was likely caused by the tubing being pulled.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discoloration and stretching or deformation.Any product that fails the visual inspection is rejected.The subject cannula would have met the specification at the time of production.The user instructions which accompany the opt970 optiflow + tracheostomy direct connection interface show in pictorial format the correct placement and fitting of the cannula and also warn: - "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death."; - "do not crush or stretch tube, to prevent loss of therapy."; - "failure to use the set-up described above can compromise performance and affect patient safety.".
 
Event or Problem Description
A distributor in canada reported via a fisher & paykel healthcare (f&p) field representative, that the tubing of an opt970 optiflow + tracheostomy direct connection interface was found broken before use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW + TRACHEOSTOMY INTERFACE
Common Device Name
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534002
MDR Report Key12866642
Report Number9611451-2021-01332
Device Sequence Number5441438
Product Code BTT
UDI-Device Identifier09420012436924
UDI-Public(01)09420012436924(10)2101530614(11)210228
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K162553
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPT970
Device Catalogue NumberOPT970
Device Lot Number2101530614
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2021
Supplement Date Received by Manufacturer01/28/2022
Initial Report FDA Received Date11/23/2021
Supplement Report FDA Received Date02/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-