| Model Number |
AA075 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Report confirmed.The device was tested and the max temperature was measured to be 151°f.The likely cause of failure could not be d etermined.It was also noted that the device failed the visual inspection.There is no evidence that any other failures during analysis were found that may have contributed to this event.Device history record #400996825 was reviewed and did not reflect any condit ions related to the current, reported malfunction.The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with the report of overheating.It was reported that there was no patient or staff impact.Repair request escalated to a product event based on reason for return and return in less than 90 days from purchase or repair.
|
| |
|