Model Number PM3562
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Erosion (1750); Unspecified Infection (1930)
Event Type  Injury  
Event Description
Related manufacturer's reference number: 2017865-2021-37600.It was reported the patient presented in the hospital.It was observed the patient had pocket erosion and an infection.The patient's pacemaker, right ventricular lead, left ventricular lead and right atrial lead were explanted on (b)(6) 2021.The patient had a new pacemaker, right ventricular lead and left ventricular implanted on (b)(6) 2021.
 
Manufacturer Narrative
The reported events of low lead impedance and no sensing were not confirmed.As received, a complete lead was returned in one piece with all four tines were found to be intact and undamaged.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspections of the lead did not find any anomalies.
 
Manufacturer Narrative
A device history record (dhr) review was performed and review of the sterilization records confirmed normal sterilization cycles for the products.The device met specifications prior to leaving abbott manufacturing facilities.The product was returned and visual inspection was normal.The cause of infection could not be traced to the device. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12869497
MDR Text Key281208878
Report Number2017865-2021-37599
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 01/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Model NumberPM3562
Device Catalogue NumberPM3562
Device Lot NumberP000126809
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/03/2021
Initial Date FDA Received11/24/2021
Supplement Dates Manufacturer Received12/20/2021
01/07/2022
01/11/2022
Supplement Dates FDA Received12/20/2021
01/10/2022
01/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/09/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUARTET
Patient Outcome(s) Required Intervention;
Patient SexMale
Patient Weight54 KG
-
-