MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG LINOX SMART T 65; ICD LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG LINOX SMART T 65; ICD LEAD Back to Search Results
Model Number 369820
Medical Device Problem Codes Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 11/13/2021
Type of Reportable Event Serious Injury
Event or Problem Description
After an implantation period of approx.85 months, oversensing with inappropriate therapies was reported.The lead was capped.
 
Additional Manufacturer Narrative
As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The analysis is thus based on the inspection of the manufacturing documents of the device as well as on the provided data.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.The analysis of the provided iegm episodes revealed artifacts on the far field and rv channel, which led to the reported oversensing as well as inappropriate shocks and icd charging cycles.In spite of the available information, no conclusion can be drawn regarding the root cause of the clinical observation.An analysis of the lead itself would be necessary to determine the root cause.Should additional information or the device itself become available for analysis, the investigation will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINOX SMART T 65
Common Device Name
ICD LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key12869738
Report Number1028232-2021-06419
Device Sequence Number3707816
Product Code NVY
UDI-Device Identifier04035479112998
UDI-Public04035479112998
Combination Product (Y/N)N
PMA/510(K) Number
P980023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2016
Device Model Number369820
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/24/2021
Supplement Date Received by Manufacturer12/03/2021
Initial Report FDA Received Date11/24/2021
Supplement Report FDA Received Date12/06/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/24/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-