Event or Problem Description
It was reported that during an initial hip arthroplasty, the pps coating deboned from the stem.Some debris was not removed from the patient as it was not easily removable.Additional information on the reported event is unavailable.
 
Additional Manufacturer Narrative
(b)(4).Source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified a spec of black debris as reported in the wound.Reported event was confirmed by review of photographs provided.A review of the device history records identified no related deviations or anomalies during manufacturing.Root cause could not be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6: component code: mechanical (g04) - stem.A tprlc 133 t1 pps so 12x144mm was returned and evaluated against the complaint.Visual inspection found metallic debris on the returned cloth and inside the poly bag that is visually consistent with the porous coating material.The porous coating appears to be thin near the collar and elbow.Blood stains and foreign debris are present on the coating and distal end of the stem.Complaint sample was evaluated and the reported event was confirmed.The additional information does not change the outcome to the previous investigation.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS SO 12X144MM T1
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12869917
Report Number0001825034-2021-03194
Device Sequence Number14958696
Product Code JDI
UDI-Device Identifier00880304491823
UDI-Public(01)00880304491823(17)301125(10)6899276
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K120030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number51-103120
Device Lot Number6899276
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2021
Supplement Date Received by Manufacturer01/28/2022
05/05/2022
Initial Report FDA Received Date11/24/2021
Supplement Report FDA Received Date02/02/2022
05/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-