| Model Number |
AU00T0 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A nurse reported during the preparation of the iol while removing "the blue lens lock" the cartridge opened and exposed iol.It seemed like the small clip holding the lens chamber closed was loose.The surgery was delayed for a few minutes but was completed with a back up lens without any enlarged incisions nor any sutures were needed.There was no patient harm and the patient was ok.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for analysis and the reported damage was not observed.Additional observations were as follows: the device is returned loose in the carton.The lens stop and the plunger is present on the device.Viscoelastic is observed in the device.The lens bay door is closed securely - it closes and opens with no issues.The lens stop can be removed with no issues.We are unable to determine the root cause for the reported complaint "cartridge opened and exposed company lens".The device was returned with the lens bay door securely closed.No damage was observed to the returned device.The returned device shows evidence of possible handling by the customer due to the presence of solution dried in the device.All product and batch history records are quality reviewed prior to product release.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|