• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG., RESECTOR SHAVER BLADE, F - SERIES - 5.5MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG., RESECTOR SHAVER BLADE, F - SERIES - 5.5MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 0375562000
Medical Device Problem Codes Break (1069); Component Missing (2306)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the device potentially broke inside the patient and was not able to be removed.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: missing part.Probable root cause: inadequate window profile, inadequate welding process (i.E.Plasma, inductive, gluing), improper material strength, inadequate size selected for the application, material brittleness , excessive force applied by the user, incorrect rfid tag.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
 
Event or Problem Description
It was reported that the device potentially broke inside the patient and was not able to be removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., RESECTOR SHAVER BLADE, F - SERIES - 5.5MM
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key12870384
Report Number0002936485-2021-00675
Device Sequence Number3547660
Product Code GFA
UDI-Device Identifier37613327061872
UDI-Public37613327061872
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0375562000
Device Lot Number21137CE2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/27/2021
Supplement Date Received by Manufacturer10/27/2021
Initial Report FDA Received Date11/24/2021
Supplement Report FDA Received Date01/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-