| Model Number |
72200568 |
| Medical Device Problem Codes |
Overheating of Device (1437); Protective Measures Problem (3015); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient was hospitalized due to lateral collateral ligament injury.During the examination of the injury, the light source 500xl xenon suddenly alerted with an e13 error, and the light went out.The procedure was successfully completed with a non-significant delay using a back-up device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.The e13 error code is a feature that alerts the user of a temperature issue and makes the devices shut down to prevent it from increasing the temperature.As a result, the device is working as intended and the issue will not contribute to serious injury.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
|
| |
|