| Model Number |
4585 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/1901
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported that the sensor "was losing connectivity." no consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned sensors were evaluated.All the sensors passed visual inspection and continuity testing.The sensors all failed the battery voltage test and did not meet the 3.4 vdc specification.During functionality testing, the lab monitor displayed an error message.
|
| |
|
Event or Problem Description
|
|
The customer reported that the sensor "was losing connectivity." no consequences or impact to patient were reported.
|
| |
|