MAUDE Adverse Event Report: TX027 TX-JACKSONVILLE-MCKNIGHT 1000CC FLEX CANISTER; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TX027 TX-JACKSONVILLE-MCKNIGHT 1000CC FLEX CANISTER; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED Back to Search Results
Model Number 65652-611
Medical Device Problem Codes Break (1069); Suction Failure (4039)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/1980
Type of Reportable Event Malfunction
Event or Problem Description
Customer reported the canisters are breaking within 4 months of use and that their shelf life is becoming shorter.Canister lost suction on high, low or both and/or cracked at the base and sometimes had decreased suction during unknown procedure.Customer has stated they use clinell wipes to clean them down and leave the canisters on high suctions 24/7 and clamp off the tubing.There was no reported adverse event to patient.No further information was provided by customer.
 
Additional Manufacturer Narrative
Supplemental report is being filed since the results of investigation are available.No lot number was provided by the customer; therefore, a review of the manufacturing device history record could not be performed.No sample was provided by the customer; therefore, an evaluation of the complaint device for deficiency of construction could not be performed.Photos were sent with the case but are not for the reported product code.Discussions were held with the team from engineering, production, and quality to determine assignable cause and corrective actions for canister that have either lost suction on high, low, or both and/or cracked at the base.The team reviewed the photos that were submitted for the sister complaint as well as the samples from the same sister complaint.It is clearly evident by the smeared and missing print that the hard canisters are being cleaned with contra-indicated cleaners spelled out in the instructions for use.The investigation has also revealed that the hard cans are left on continuous suction 24/7.This practice is contra-indicated by the instructions for use.Solvents attack the areas of highest stress concentration which is around the bottom edge of the canister and at the gate (center of the bottom) making it weak.The continuous high suction will cause the hard canister to break at the weak spots.The combination of contra-indicated cleaners and the continuous high suction will create conditions for canister defects.The instructions for use detail information regarding continuous suction and what type of cleaners to use on the hard canisters, as stated: hardware warnings: 1.Cleaning agents that contain quatemary ammonium salts, strong alkalines or chlorinated benezene (phenols) based hydrocarbons may cause chemical degradation of the material.This could lead to a failure of the products to perform as intended.Warnings: 1.This product is not designed to be kept under vacuum when not in use.Prolonged static vacuum may increase the risk of cracking or implosion.The assignable cause is customer misuse.The damage to the canisters is caused by using contra-indicated cleaners which weaken the polycarbonate followed by continuous high suction 24/7.Trending of past complaint reports indicate that this is the 1st reported incident of canister cracking at bottom for product code 65652-611 in the past three calendar years.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1000CC FLEX CANISTER
Common Device Name
APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED
Manufacturer (Section D)
TX027 TX-JACKSONVILLE-MCKNIGHT
200 mcknight st.
jacksonville TX 75766
Manufacturer (Section G)
TX027 TX-JACKSONVILLE-MCKNIGHT
200 mcknight st.
jacksonville TX 75766
Manufacturer Contact
patricia tucker
1500 waukegan rd
waukegan, IL 60085
8478874151
MDR Report Key12872601
Report Number1423537-2021-00691
Device Sequence Number19274070
Product Code GCX
UDI-Device Identifier10885380042638
UDI-Public10885380042638
Combination Product (Y/N)N
Initial Reporter CountryNZ
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number65652-611
Device Catalogue Number65652-611
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/03/2021
Supplement Date Received by Manufacturer11/03/2021
Initial Report FDA Received Date11/24/2021
Supplement Report FDA Received Date12/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-