• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MSD DEGGENDORF MFG CLYDESDALE PTC¿ SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MSD DEGGENDORF MFG CLYDESDALE PTC¿ SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, Back to Search Results
Model Number 4986850
Medical Device Problem Codes Break (1069); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2021
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding an event happened during intra-op for a patient undergoing l2-l5 olif.The pre-op diagnosis was mentioned as "degenerative lumbar scoliosis".It was reported that, in the case of l2-5 olif, after the olif was performed at l4 / 5 and l3 / 4, the cage did not enter at all from the middle during the olif of l2 / 3.The cage could no longer be tapped and inserted, and the inserter could not be attached because the thread part of the cage was broken.They used a remover, but the cage did not move at all.Finally they grabbed the cage with luer and managed to remove it.They replaced it with a new cage, but the second cage had the same event and was removed by the same means.No fragments of broken cages remained in patient.There was a delay of more than 60 minutes reported as a result.No further complications reported regarding the event.
 
Additional Manufacturer Narrative
Evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
H3: product analysis of part#: (b)(4), lot#: h5688195.Analysis summary: visual and microscopic examination identified, the threads of the inserter interface to be stripped off, with evidence of deformation on the face of the implant or hole diameter.The nose of the implant as witness marks in the center.The above observations are consistent with significant impact, during attempted assembly.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank, because the information is currently, unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLYDESDALE PTC¿ SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
Manufacturer (Section D)
MSD DEGGENDORF MFG
wertstrasse 17
deggendorf PR 94469
GM  94469
Manufacturer (Section G)
MSD DEGGENDORF MFG
wertstrasse 17
deggendorf PR 94469
GM   94469
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key12874325
Report Number1030489-2021-01457
Device Sequence Number14612190
Product Code MAX
UDI-Device Identifier00643169190528
UDI-Public00643169190528
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133205
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4986850
Device Catalogue Number4986850
Device Lot NumberH5688195
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2021
Supplement Date Received by Manufacturer04/29/2022
Initial Report FDA Received Date11/25/2021
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
Patient Weight60 KG
-
-