SMITH & NEPHEW, INC. GII CR DEEP FLX ISR TR S5-6 11; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
|
Back to Search Results |
|
| Model Number |
71430447 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a genesis ii cruciate retaining deep flexion insert trial size 5-6 9mm (case: (b)(4)), a genesis ii cruciate retaining deep flexion insert trial size 5-6 11m (case: (b)(4)) and a genesis ii cruciate retaining deep flexion insert trial size7-8 11m (case: (b)(4)) need replacing due to wear and tear, plastic burs, gouges, cracked ends, among others.As this was noticed upon field inspection, there was not patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, intended for use in treatment, was returned for evaluation.The visual inspection confirms the device has multiple burrs, deep gouges, scratches and there are cracks in the plastic.This device shows significant signs of wear/usage.A review of complaint history did not reveal additional complaints for the listed batch.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|