• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105NNPKNA NOVO NORDISK PINK; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105NNPKNA NOVO NORDISK PINK; SYRINGE, PISTON Back to Search Results
Model Number MMT-105NNPKNA
Device Problem Computer Software Problem (1112)
Patient Problem Insufficient Information (4580)
Event Date 11/12/2021
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Information received by medtronic indicated that the insulin pen had a dose log inaccuracy and doses were not randomly not logged.No harm requiring medical intervention was reported.The insulin pen will be returned for analysis.
 
Manufacturer Narrative
Serial number: (b)(6).Software version: 3.8.4.Color: pink.Battery life remaining: <4 months.Customer reports: doses are not being registered on time or are only being registered approximately half an hour later and sometimes, will not show the amount in the dose.Per visual inspection: no physical damage to injection foot or inpen was noted.Unit paired successfully to commercial app.The following test values were dialed and dosed: 10 test values of 4.0u dialed and dosed.All values displayed accurately in the logbook.All doses were written to the app logbook within 1 minute.Inpen received with leadscrew halfway of the travel.Re-wound screw.No drag was observed, the screw was not bent, advanced when dosage knob was pressed dialing a dosage and retracted appropriately.No resistance was observed when dosing without a cartridge installed.The screw advanced every time 30.0u was dialed and dosed until the screw reached max extension.In conclusion: inpen received working as design.No malfunctions noted during testing that could affect insulin delivery.The patient complaint of dose log inaccuracies could not be confirmed.Inpen will be sent to san diego for further testing.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNPKNA NOVO NORDISK PINK
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
tricha miles
1223 world trade drive, 100
san diego, CA 92128
7635140379
MDR Report Key12874679
MDR Text Key282490911
Report Number3012822846-2021-00829
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-105NNPKNA
Device Catalogue NumberMMT-105NNPKNA
Device Lot NumberB93CT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/12/2021
Initial Date FDA Received11/25/2021
Supplement Dates Manufacturer Received01/19/2022
Supplement Dates FDA Received01/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age3 YR
Patient SexFemale
Patient Weight40 KG
-
-