• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715K
Device Problems Excess Flow or Over-Infusion (1311); Appropriate Term/Code Not Available (3191)
Patient Problems Hyperglycemia (1905); Hypoglycemia (1912); Loss of consciousness (2418)
Event Date 11/14/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that they experienced a hyperglycemia and hypoglycemia.Customer blood glucose level at the time of incident was 400 mg/dl.Customer low blood glucose level was treated via glucose.Customer was used insulin pump system within 48 hours of reported event.Customer alleged that the pump was over delivering.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Retainer ring = black on (b)(6) 2021, the customer alleged the pump over delivery and hospitalized for low and high bg.The test p-cap locks properly in place in the reservoir compartment noted.Pump received with pillowing keypad overlay during the visual inspection.Thus and carelink software was utilized and downloaded trace/history files properly.In further full review of the pump history on the event date of (b)(6) 2021, found two bolus delivery that the customer manually programmed and delivered at 12:47:44 of 4.6u and 21:01:04 of 3.4u.Pump passed the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and dat at 0.08700 inches.Pump was cut open to perform visual inspection and found no moisture or component damage on the electronics, force sensor and motor assembly noted.The force sensor offset measured (23.0 mv).The motor was tested outside of the device on the ngp stb3 and passed.In summary, pump passed all required testing.Unable to verify customer complaint for low and high bg.Customer alleged for the pump over delivery was not confirmed.The force sensor is within specification and the motor functioning properly.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715K 630G BLACK MG
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key12876281
MDR Text Key282058600
Report Number2032227-2021-222732
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000763000316631
UDI-Public(01)000000763000316631
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-1715K
Device Catalogue NumberMMT-1715K
Device Lot NumberHG4677K
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received03/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/13/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
FRN-UNK-RSVR, UNOMED INF SET
Patient Outcome(s) Other;
Patient Age69 YR
Patient SexFemale
Patient Weight84 KG
-
-