Report source.Country: (b)(6).Device evaluated by mfr: neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
|
Information was received from a healthcare provider (hcp) via manufacturer representative regarding a patient having l3/l4 material ablation for an indication of infection.It was reported that during the locking of the left screw at the bottom l4, the screw extension broke before the nut, causing the screw to fall out of the pedicle.The extension failure pushed the screw laterally and it caused a pedicle fracture.The surgeon had to remove all the screws and no more implant possible on this level.There was no surgical delay reported.There were no further complications reported.
|
|
Device evaluation summary: ref: (b)(4), lot: 20f0696 visual control.Analysis was performed with sme: the instrument for tightening the nut would not have been used in the right axis.When it comes to tighten and lock the screw, the lever arm caused significant efforts on the pedicle which ended up breaking both the screw and the pedicle.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|