• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDICREA INTERNATIONAL SA PASS LP; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDICREA INTERNATIONAL SA PASS LP; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number B02216545
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 10/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Report source.Country: (b)(6).Device evaluated by mfr: neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider (hcp) via manufacturer representative regarding a patient having l3/l4 material ablation for an indication of infection.It was reported that during the locking of the left screw at the bottom l4, the screw extension broke before the nut, causing the screw to fall out of the pedicle.The extension failure pushed the screw laterally and it caused a pedicle fracture.The surgeon had to remove all the screws and no more implant possible on this level.There was no surgical delay reported.There were no further complications reported.
 
Additional Manufacturer Narrative
Device evaluation summary: ref: (b)(4), lot: 20f0696 visual control.Analysis was performed with sme: the instrument for tightening the nut would not have been used in the right axis.When it comes to tighten and lock the screw, the lever arm caused significant efforts on the pedicle which ended up breaking both the screw and the pedicle.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASS LP
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MEDICREA INTERNATIONAL SA
5389 rte de strasbourg
rillieux la pape,fr 69140
FR  69140
Manufacturer (Section G)
MEDICREA INTERNATIONAL SA
5389 rte de strasbourg
rillieux la pape,fr 69140
FR   69140
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key12876693
Report Number1000432246-2021-00013
Device Sequence Number11894267
Product Code NKB
UDI-Device Identifier03613720177586
UDI-Public03613720177586
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K141398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberB02216545
Device Catalogue NumberB02216545
Device Lot Number20F0696
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2021
Supplement Date Received by Manufacturer02/07/2022
Initial Report FDA Received Date11/26/2021
Supplement Report FDA Received Date02/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-