• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ELEKTA INC MOSAIQ; ACCELERATOR, LINEAR, MEDICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ELEKTA INC MOSAIQ; ACCELERATOR, LINEAR, MEDICAL Back to Search Results
Model Number MOSAIQ
Device Problems Computer Software Problem (1112); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/28/2021
Event Type  malfunction  
Manufacturer Narrative
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
 
Event Description
The customer reported that the linac delivered a partial field and then it stopped for some issue, but mosaiq recorded the field as completed.
 
Manufacturer Narrative
Section h6 updated coding.Section h10 additional information.The investigation was completed by conducting a thorough evaluation of the product and the reported information.It is possible for the displayed delivery mu to rise to the prescribed dose following an abnormal dose diff termination.The integrity software will show an error and report the abnormal termination to the record & verify system (mosaiq will display an error).In the issue reported mosaiq was in the process of recording the beam records when a machine error occurred.The user was presented with an abnormal termination form.The user verified the actual treatment and corrected any incomplete treatment.There was no patient mistreatment.Mosaiq did not have any malfunction and was working as designed and intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Type of Device
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
ELEKTA INC
100 mathilda place
5th floor
sunnyvale CA 94086
Manufacturer (Section G)
ELEKTA INC
400 perimeter center terrace
suite 50
atlanta GA 30346
Manufacturer Contact
pms
cornerstone
london road
crawley, west sussex RH10 -9BL
UK   RH10 9BL
MDR Report Key12877216
MDR Text Key281278740
Report Number2950347-2021-00021
Device Sequence Number1
Product Code IYE
UDI-Device Identifier00858164002091
UDI-Public00858164002091
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K141572
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberMOSAIQ
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/29/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-