Model Number S-45-120-120-P6
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a 90% stenosed, moderately calcified and moderately tortuous vessel in the superficial femoral artery (sfa).The vessel diameter was 4.5mm and atherectomy was not used in the procedure.Pre-dilatation was performed using a 5mmx120mm balloon with a pressure of 8 atmospheres (atms) for 5min.The 4.5x120mm supera self-expanding stent system (sess) was attempted to be released but could not be separated from the delivery sheath.The stent and the delivery sheath were withdrawn together from the anatomy under fluoroscopy.Another supera sess was used to complete the procedure.There was no adverse patient effect and there was no reported clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.It is possible that interaction with the 90% stenosed, moderately calcified and moderately tortuous anatomy resulted in the ratchet being unable to properly engage and release the stent from the delivery system thus resulting in the reported activation failure/ deployment failure; however this cannot be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key12878257
Report Number2024168-2021-10820
Device Sequence Number4472440
Product Code NIP
UDI-Device Identifier08717648226007
UDI-Public08717648226007
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Model NumberS-45-120-120-P6
Device Catalogue NumberS-45-120-120-P6
Device Lot Number0120761
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/08/2021
Supplement Date Received by Manufacturer11/29/2021
Initial Report FDA Received Date11/26/2021
Supplement Report FDA Received Date12/14/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-