A healthcare professional reported that during a cataract/ vitrectomy combination surgery, an ophthalmic operating console had with no aspiration.The surgeon changed the anterior vitrectomy probe, and then operated after the connections and priming, the irrigation worked along with cutting, however there was no aspiration.The system was primed again but the issues were persisted.Movement of the flow of fluid or air bubbles in the aspiration tube was checked for but there was no issue found.There was no system occlusion tone heard.The second phacoemulsification console connected with probe and it worked without any issue.The vitrectomy was completed with the second console.The cortical mass was cleaned up and the aspiration of the healon was completed by first console with the irrigation/aspiration probe.The required procedure was completed on the same day.There was an intraocular bleed which tracked into the vitreous.There was a delay in completing the surgery as due to repeated priming and a change of console.The patient is being followed up for the vitreous hemorrhage.
|
A sample was not received at the manufacturing site for evaluation.The condition of the product could not be verified.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.A sample was not received at the manufacturing site and no lot information was provided; therefore, the root cause for the customer complaint issue cannot be determined.The cause of the reported event cannot be determined with the information obtained; therefore, specific action with regards to this complaint cannot be taken.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|