ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number ASKU
Device Problem Decrease in Suction (1146)
Patient Problem Vitreous Hemorrhage (2143)
Event Date 10/22/2021
Event Type  Injury  
Event Description
A healthcare professional reported that during a cataract/ vitrectomy combination surgery, an ophthalmic operating console had with no aspiration.The surgeon changed the anterior vitrectomy probe, and then operated after the connections and priming, the irrigation worked along with cutting, however there was no aspiration.The system was primed again but the issues were persisted.Movement of the flow of fluid or air bubbles in the aspiration tube was checked for but there was no issue found.There was no system occlusion tone heard.The second phacoemulsification console connected with probe and it worked without any issue.The vitrectomy was completed with the second console.The cortical mass was cleaned up and the aspiration of the healon was completed by first console with the irrigation/aspiration probe.The required procedure was completed on the same day.There was an intraocular bleed which tracked into the vitreous.There was a delay in completing the surgery as due to repeated priming and a change of console.The patient is being followed up for the vitreous hemorrhage.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation.The condition of the product could not be verified.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.A sample was not received at the manufacturing site and no lot information was provided; therefore, the root cause for the customer complaint issue cannot be determined.The cause of the reported event cannot be determined with the information obtained; therefore, specific action with regards to this complaint cannot be taken.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key12879857
MDR Text Key281319317
Report Number1644019-2021-00756
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/27/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-