• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD ENDO CLIP; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD ENDO CLIP; CLIP, IMPLANTABLE Back to Search Results
Model Number 176615
Medical Device Problem Codes Difficult to Open or Close (2921); Mechanics Altered (2984)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a liver surgery, the clip was stuck in the jaws of the device during half of firing and the firing could no longer be carried out.The jaws was not able to close.A new device was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic right hepatectomy, the clip was stuck in the jaws of the device during half of firing and the firing could no longer be carried out.The jaws were not able to close.A new device was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the handle was broken and out position, in a manner consistent of cycling through the safety interlock.It was reported that the clips did not load properly into the jaws as expected and the jaws of the reload did not close at all.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur when an attempt is made to fire the instrument without first loading a clip by using the trigger.The safety interlock feature is designed to prevent the jaws from closing on a vessel in the absence of a loaded clip.If an attempt is made to forcibly fire the instrument while engaged in interlock, the lugs are designed to break as noted and the interlock feature continues to function as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: before attempting to squeeze the handle, be sure to first load a clip into the instrument jaws.If the jaw is empty, the instrument will not close.Always check to see if a clip is seated in the jaws prior to firing.When squeezing the handle, remove the finger from the trigger.Do not simultaneously pull the trigger and squeeze the handle as this may result in the firing of an unformed clip or jamming of the instrument.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key12882135
Report Number9612501-2021-01958
Device Sequence Number14459058
Product Code FZP
UDI-Device Identifier10884521057807
UDI-Public10884521057807
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K143644
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model Number176615
Device Catalogue Number176615
Device Lot NumberJ0C1136NY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2021
Supplement Date Received by Manufacturer11/25/2021
02/16/2022
Initial Report FDA Received Date11/29/2021
Supplement Report FDA Received Date12/13/2021
02/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-