C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER; TEMPERATURE SENSING FOLEY CATHETER Back to Search Results
Model Number 119108
Medical Device Problem Codes Partial Blockage (1065); Inaccurate Flow Rate (1249)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the all-silicone temperature sensing foley catheter was removed and water was injected since there was no urine flow, but the situation did not change.Per additional information received via email from ibc on (b)(6) 2021, it was confirmed that the blockage in the catheter was unknown and there was no urine flow.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the all-silicone temperature sensing foley catheter was removed and water was injected since there was no urine flow, but the situation did not change.Per additional information received via email from ibc on 10nov2021, it was confirmed that the blockage in the catheter was unknown and there was no urine flow.
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing related.1 sample was confirmed to exhibit the reported failure.The device had not met specifications.The product was used for treatment.The product had caused the reported failure.Visual evaluation of the returned sample 1 used silicone foley.Visual inspection of the sample noted 1 three-way temp sensing catheter the drainage lumen was flushed with 3.5ml of methylene blue solution (3 drops 1% aq methylene blue per 100ml of distilled water) and the solution did not advance through the funnel observed a blockage on the return sample.This is out of specification per inspection procedure which states, "funnel must not have distortions or bent at shaft connection with the funnel." the device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labelling could not have prevented the reported failure.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER
Common Device Name
TEMPERATURE SENSING FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12882664
Report Number1018233-2021-07551
Device Sequence Number12035865
Product Code EZL
UDI-Device Identifier00801741045998
UDI-Public(01)00801741045998
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/11/2025
Device Model Number119108
Device Catalogue Number119108
Device Lot NumberNGEW0060
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/08/2021
Supplement Date Received by Manufacturer05/05/2022
Initial Report FDA Received Date11/29/2021
Supplement Report FDA Received Date05/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-