• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH 30 SILICONE PIP IMPLANT (STERILE PACKED); PROSTHESIS, FINGER, CONSTRAINED, POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH 30 SILICONE PIP IMPLANT (STERILE PACKED); PROSTHESIS, FINGER, CONSTRAINED, POLYMER Back to Search Results
Model Number PIP-30
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Joint Dislocation (2374)
Date of Event 10/30/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The customer reported the following issue: "silicone implant broke off, causing finger to dislocate.The patient required a revision surgery.".
 
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device disposition is unknown.
 
Event or Problem Description
The customer reported the following issue: "silicone implant broke off, causing finger to dislocate.The patient required a revision surgery.".
 
Additional Manufacturer Narrative
Please note corrections to sections d9/h3 the device was returned, and h6 (method and results codes).The reported event could be confirmed, since the device was returned, and is indeed broken in two.One of the arms of the silicone implant is broken, most probably because of the wear of the material after multiple cycles of flexion-extension.More detailed information about the complaint event such as some patient info (activity levels), x-rays, as well as the implantation date must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30 SILICONE PIP IMPLANT (STERILE PACKED)
Common Device Name
PROSTHESIS, FINGER, CONSTRAINED, POLYMER
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
marilyne chaumont
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key12882905
Report Number0008031020-2021-00477
Device Sequence Number4827858
Product Code KYJ
UDI-Device Identifier00886385021553
UDI-Public00886385021553
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K931588
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberPIP-30
Device Catalogue NumberPIP30
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/02/2021
Supplement Date Received by Manufacturer03/16/2022
Initial Report FDA Received Date11/29/2021
Supplement Report FDA Received Date04/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age72 YR
Patient SexMale
-
-