BOSTON SCIENTIFIC CORPORATION MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7590
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 85% stenosed target lesion was located in the moderately tortuous and moderately calcified mid left anterior descending artery.A 1.50mm x 20mm maverick balloon catheter was advanced for dilatation.However, during inflation, the pressure did not rise to more than 10 atmospheres.Upon checking, a small hole in the marker was found.The device was completely removed and the procedure was completed with another of same device.No patient complications were reported and the patient condition was stable.
 
Additional Manufacturer Narrative
Returned product consisted of a maverick 2mr balloon catheter.The device was visually and microscopically examined.There were numerous kinks to the hypotube of the device.There was contrast present in the inflation lumen and balloon.The balloon was loosely folded, and the tip of the device was damaged.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There was a pinhole in the balloon located 15mm from the tip of the device.The device failed to inflate to rated burst pressure.Product analysis confirmed the reported event, as during functional testing the device was found to have a pinhole damage to the balloon.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 85% stenosed target lesion was located in the moderately tortuous and moderately calcified mid left anterior descending artery.A 1.50mm x 20mm maverick balloon catheter was advanced for dilatation.However, during inflation, the pressure did not rise to more than 10 atmospheres.Upon checking, a small hole in the marker was found.The device was completely removed and the procedure was completed with another of same device.No patient complications were reported and the patient condition was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12884729
Report Number2134265-2021-14952
Device Sequence Number3900292
Product Code LOX
UDI-Device Identifier08714729370154
UDI-Public08714729370154
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/29/2024
Device Model Number7590
Device Catalogue Number7590
Device Lot Number0026717358
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/22/2021
Supplement Date Received by Manufacturer01/10/2022
Initial Report FDA Received Date11/29/2021
Supplement Report FDA Received Date01/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-